ToshibaメーカーTIC-LF414Cの使用説明書/サービス説明書
ページ先へ移動 of 16
TIC-LF414C Introduction The electromagnetic flowmeter uses Faraday’s Law of electromagnetic induction to measure the process flow. The device consists of two units: a detector, through which the flu.
TIC-LF414C 2 Power consumption: 17W(27VA) or less 19W(29VA) or less (with PROFIBUS) Conformance to European Co mmunity Directives: EMC directive 89/336/EEC The low voltage 93/68/EEC PED 97/23/EC (Note 1) Note : See Table 1 for detail.
TIC-LF414C 3 Digital input DI (opt.) Signal type: 20 to 30Vdc voltage signal Input resistance: 2.7k Ω Number of inputs: one point DI function — One of the following functions can be assigned to the optional DI signal. Range switching — Selects either the higher or lower range in the unidirectional or bidirectional 2-range setting.
TIC-LF414C 4 Conditions when power fails: Parameter setting values are stored in non–volatile memory and the values will be restored when the power r eturns to normal condition.
TIC-LF414C 5 Installation Dimensions Unit : inch( mm) L1 Φ D1 (L2 ) 2. 5 2 ( 64 ) 5.63( 143) 1.42 ( 36) 8.86 ( 225) 5.16( 131) ANSI class 150 and class 300 dimensions: Meter size (inch) L1 (inch) L2 (inch) D1 (inch) Weight (lb) 1/2 2.76 9.92 1.93 approx.
TIC-LF414C 6 Unit : inch(mm) (L2 ) L1 8.86( 225) 5.63 ( 143) 1.42( 36) 5.16( 131 ) Φ D1 Note1: Eye bolts are provided at the top fo r flow mete rs sized 8"(200mm). Note2: 1 inch = 25.4mm ANSI class 150 and class 300 dimensions Meter size (inch) L1 (inch) L2 (inch) D1 (inch) Weight (lbs) 1-1/2 3.
TIC-LF414C 7 Unit : inch(mm) R ( P T ) 1/2Male S c r e w ( T w o Pl a c e s) 3.46( 88) 1.57 ( 40 ) 4.96 ( 126 ) 3.58 ( 91 ) 1.38( 35) (L2 ) Φ 1.93( Φ 49) 2.52( 64 ) L1 Note1: Cable glands are not provide d for LF414 of FM and CSA appr oved type. Refer to the part of Cable connection port at detector.
TIC-LF414C 8 Unit : inch(mm) 3.46(88) 1.57 ( 40 ) 4.96(126) 1.38(35) 3.58(91) R(PT)½ Mal e Screw (T wo places) (L2) Φ 2.60( Φ 66) 3.43 ( 87 ) L1 Note1: Cable glands are not provide d for LF414 of FM and CSA appr oved type. Refer to the part Cable c onnection port at detector.
TIC-LF414C 9 Unit : inch(mm) 3.46( 88) 1 .57 ( 4 0 ) 4.9 6( 126) 1. 38 ( 3 5 ) 3.58(91) R( PT )½M a le Sc r e w ( T w o p l a c e s ) (L 2) L1 Φ D1 Note1: Cable glands are not provide d for LF414 of FM and CSA appr oved type. Refer to the part of Cable connection port at detector.
TIC-LF414C 10 矢視 A 信号ケ ー ブ ル グラ ン ド 電源ケ ー ブ ル グ ラ ン ド 封止栓 励磁 ケ ー ブ ル グ ラ ン ド 電流出力ケ ー ブ ル グ ラ ン ド 接地端子 (M 4 ) 取付金具 4 - φ 11 取付金具 82 (3 6 ) 1 5 6 1 3 1 51 178 20 0 1 51 74 ± 0 .
TIC-LF414C 11 ・ Separate type LF410/LF612 and LF414/LF612F flowmeters T erminal board Connected detecto r IV wire 5.5 mm 2 or more Power switch (External double- pole power switch ) Thick walled ste.
TIC-LF414C 12 Wiring Precautions (1) Explosion proof type flow meters are not provided cable glands. Refer to the part Cable connection port at detector and converter. (2) Connect the grounding wire (IV wire 5.5mm² or more) to a good earth ground (100 Ω or less ground resistance).
TIC-LF414C 13 Meter Size To select the meter size: See Table 3 to 4 and find meter sizes within the velocity of 0.3 to 32.8 ft/s (0.1 to 10m/s) for a specified full-scale (measuring range high limit) flow. Select one that has its full-scale velocity between 3.
TIC-LF414C 14 Table 6. Specification Code (Waf er type detector LF410 Series) Detector ca tegory Model Specific ation Code Normal type Ex. type 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Description Gr.
TIC-LF414C 15 Table 7. Specification Code for converter s Model Specific ation Code 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Contents L F 6 1 Elec tromagnetic flowmeter c onverter LF610 type LF612 type 0 Comb.
TIC-LF414C TIC-LF414C Specifications are subject to change without notice. Printed in Japan 2008-6 (TDOC) © TOSHIBA Corporation 2008 All Rights Reserved. http://www.toshiba.com/ind / Misuse of this product can result in damages to property or hum an injury.
デバイスToshiba TIC-LF414Cの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Toshiba TIC-LF414Cをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはToshiba TIC-LF414Cの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Toshiba TIC-LF414Cの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Toshiba TIC-LF414Cで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Toshiba TIC-LF414Cを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はToshiba TIC-LF414Cの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Toshiba TIC-LF414Cに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちToshiba TIC-LF414Cデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。